Tektronix AWG610 Manuel d'utilisateur Page 385

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 466
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 384
Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-25

DMM
50 W BNCĂcoaxial cable
InputĂconnector
+ĂBNCĆtoĆdual banana adapter
+Ă50 W BNCĂcoaxialĂcable
AWG610 Arbitrary Waveform Generator
CH1ĂoutputĂconnector
+ĂSMA(Ma)ĆBNC(Fe) adapter
+ĂBNC 50 W Ăprecision terminator
+Ă50 WĂBNC coaxial cable
Figure B-9: Amplitude accuracy initial test hookup
2. Set the DMM controls as follows:
Mode . ......................... VDC
Range ......................... Auto
Input .......................... Front
3. Follow the substeps below to set the AWG610 Arbitrary Waveform
Generator controls and to select the sequence file:
a. Push UTILITY (front-panel)System (bottom)Factory Reset
(side)OK (side).
b. Push SETUP (front-panel)Run Mode (bottom)Enhanced (side).
The AWG610 Arbitrary Waveform Generator is set to enhanced mode.
c. Load the AMP1.SEQ file.
Refer to Loading Files on page B–10 for file loading procedures.
4. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connector are on.
Vue de la page 384
1 2 ... 380 381 382 383 384 385 386 387 388 389 390 ... 465 466

Commentaires sur ces manuels

Pas de commentaire