Tektronix AWG610 Manuel d'utilisateur Page 415

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 466
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 414
Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual

3. Follow the substeps below to set the AWG610 Arbitrary Waveform
Generator controls:
a. Push UTILITY (front-panel)System (bottom)Factory Reset
(side)OK (side).
b. Load the MODE.WFM file.
Refer to Loading Files on page B–10 for file loading procedures.
c. Push HORIZONTAL MENU (front-panel)Clock (side).
d. Push 2, 0, 0 and M (SHIFT+7) keys in this order or turn the general pur-
pose knob to set the internal clock frequency to 200 MHz.
4. Push the RUN button.
The LED above the RUN button is on.
5. Verify that the clock signal amplitude is equal to or larger than 0.5 V, and the
clock signal period is 20 ns (50 MHz).
6. Disconnect the cable from the 1/4 Clock connector.
Vue de la page 414
1 2 ... 410 411 412 413 414 415 416 417 418 419 420 ... 465 466

Commentaires sur ces manuels

Pas de commentaire