Performance Verification
4–8
TDS3000 Series Service Manual
18. Push the –more– screen button until you can select the Pk-Pk measure-
ment.
19. Set the vertical SCALE to 100 mV/div.
20. Set the output frequency of the leveled sine wave generator to 50 kHz.
21. Set the output amplitude of the leveled sine wave generator so the peak-to-
peak measurement is between 599 mV and 601 mV.
22. Set the horizontal SCALE to 10 ns/div.
23. Set the output frequency of the leveled sine wave generator to the frequency
shown in the table below.
Oscilloscope model Frequency
TDS3012, TDS3014 100 MHz
TDS3032, TDS3034 300 MHz
TDS3052, TDS3054 500 MHz
24. Verify that the peak-to-peak measurement is 425 mV.
25. Push the waveform off
button.
26. Repeat steps 11 through 25 for each channel of the oscilloscope (not
including the external trigger input on some models).