Application Examples
TDS1000/2000-Series Digital Oscilloscope User Manual
49
Measuring Ring Amplitude
You measured the ring frequency in the previous example. Now you
want to measure t he amplitude of the ringing. To measure the
amplitude, follow these steps:
1. Push the CURSOR button to see the Cursor Menu.
2. Push the Type option button and select Voltage.
3. Push the Source option button and select CH1.
4. Turn the CURSOR 1 knob to place a cursor on the highest peak
of the ring.
5. Turn the CURSOR 2 knob to place a cursor on the lowest point
of the ring.
You can see the following measurements in the Cursor Menu:
H The delta voltage (peak-to-peak voltage of the ringing)
H The voltage at Cursor 1
H The voltage at Cursor 2