Tektronix TPG20 Manuel d'utilisateur Page 17

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 102
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 16
Performance Check and Calibration Procedures
TPG20 Service
1–5
Performance Check
1. Visual Inspection
a. Connect the TPG20 to a power source, and turn the power on.
b. Check – the front panel LCD contrast. The readout should be easily
visible under normal lighting conditions.
c. Check – all front panel controls, indicators and rear panel connectors
for functionality, appearance, and mechanical integrity.
2. Subcarrier Accuracy
NOTE. After initial delivery or long storage allow a two–hour warm up to re-age
the crystal. Thereafter, 30 minutes warm up is sufficient.
NTSC 3.579545 MHz 2 Hz
a. Set TPG20:
FORMAT – 525/60 COMPOSITE NTSC/D2
Groups; FLAT FIELDS; BLACK 525 NTSC/D2
(See Figure 1–1)
525/60 COMPOSITE
NTSC / D2
Groups:
BARS
BLACK
525 NTSC / D2 00
Search for
Format
Search for
Group
Search for
Pattern
SELECT
or
PATTERN
SELECT
or
PATTERN
SELECT
FORMAT
     
Step 1
Step 2
Step 3
Step 4
Figure 1–1: Basic steps to set the TPG20 output pattern
Vue de la page 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 101 102

Commentaires sur ces manuels

Pas de commentaire