
Removal and Installation Procedures
6–18
TDS 684A, TDS 744A, & TDS 784A Service Manual
5. Reinstallation: Do in reverse steps 3 and 4 to reinstall the line cord and then
the line fuse.
Line Fuse
Fuse Cap
1 Unplug
2 Rotate
3 Pull
Figure 6–5: Line Fuse and Line Cord Removal
See Rear Cover and Cabinet procedure on page 6–18.
1. Assemble equipment and locate modules to be removed:
a. Have handy a screwdriver with a size T-20 TorxR
tip (Items 1 and 3).
b. Make sure the oscilloscope’s front cover is installed; if it’s not, install it
by snapping its edges over the trim ring.
c. Locate the rear cover and cabinet in the locator diagram External
Modules, Figure 6–1.
2. Orient the oscilloscope: Set the oscilloscope so its face is down with its front
cover on the work surface and its bottom facing you.
EMI Gaskets
Rear Cover and Cabinet
Commentaires sur ces manuels