Tektronix AFG3000 Manuel d'utilisateur Page 105

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 230
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 104
Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-31
*OPC
This command generates the operation complete message by setting bit 0 in the
Standard Event Status Register (SESR) when all pending commands that generate
an OPC message are complete.
The query command places the ASCII character “1” into the output queue when all
such OPC commands are complete.
Group
Synchronization
Syntax
*OPC
*OPC?
Arguments
None
Returns
<execution complete>::=1
where “1” indicates that all pending operations are complete.
Examples
*OPC?
might return 1 to indicate that all pending OPC operations are finished.
*OPT?
This query-only command returns a list of the options installed in your arbi-
trary/function generator.
Group
System
Syntax
*OPT?
Arguments
None
Returns
<OPT>[,<OPT>[,<OPT>[,<OPT>]]]
Examples
*OPT?
might return 0, which indicates no option is installed in the instrument.
Vue de la page 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 229 230

Commentaires sur ces manuels

Pas de commentaire