Syntax and Commands
3-36 AFG3000 Series Arbitrary/Function Generators Reference Manual
*RST (No Query Form)
This command resets the instrument to the factory default settings. This command
is equivalent to pushing the Default button on the front panel. The default values
are listed in Default Settings on page D-1.
Group
System
Syntax
*RST
Arguments
None
Examples
*RST
resets the arbitrary/function generator settings to the factory defaults.
*SAV (No Query Form)
This command stores the current settings of the arbitrary/function generator to a
specified setup memory location.
A setup memory location numbered 0 (last setup memory) is automatically over-
written by the setups when you power off the instrument.
If a specified numbered setup memory is locked, this command causes an error.
Group
Memory
Related Commands
*RCL
Syntax
*SAV {0|1|2|3|4}
Arguments
0, 1, 2, 3, or 4 specifies the location of setup memory.
Examples
*SAV 2
saves the current instrument state in the memory location 2.