Tektronix AFG3000 Manuel d'utilisateur Page 162

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 230
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 161
Syntax and Commands
3-88 AFG3000 Series Arbitrary/Function Generators Reference Manual
[SOURce[1|2]]:VOLTage:UNIT
This command sets or queries the units of output amplitude for the specified
channel. This command does not affect the offset, High level, or Low level of
output. The setting of this command is not affected by the units setting of
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate][:AMPLitude] command.
If your instrument is a dual-channel model and the
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe] command is set to ON, then the
units of the other channel are set the same.
Group
Source
Related Commands
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe]
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
Syntax
[SOURce[1|2]]:VOLTage:UNIT {VPP|VRMS|DBM}
[SOURce[1|2]]:VOLTage:UNIT?
Arguments
VPP sets the units of the output voltage to V
p-p
.
VRMS sets the units of the output voltage to V
rms
.
DBM sets the units of the output voltage to dBm. You cannot specify DBM if the
load impedance is set to infinite.
Returns
VPP|VRMS|DBM
Examples
SOURce1:VOLTage:UNIT VPP
sets the voltage units to V
p-p
.
Vrms
Vpp
22
----------
(sin)=
dBm 10 log
10
×
P
0.001
---------------
⎝⎠
⎛⎞
=
P
Vrms
2
R L
---------------=
RL load impedance Vrms
Vpp
23
----------
(triangle)=
Vue de la page 161
1 2 ... 157 158 159 160 161 162 163 164 165 166 167 ... 229 230

Commentaires sur ces manuels

Pas de commentaire