Performance Tests
TDS 500C, TDS 600B & TDS 700C Performance Verification and Specifications
1–67
Option 05 Video Trigger Checks
Equipment
required
PAL signal source (Item 14)
60 Hz. sine wave generator (Item 15)
Pulse generator (Item 16)
Two 75 cables (Item 17)
Two 75 terminators (Item 18)
One BNC T connector (Item 7)
50 cable (Item 5)
50 terminator (Item 4)
Prerequisites See page 1–15. These prerequisites include running the signal path
compensation routine.
1. Set up digitizing oscilloscope to factory defaults by completing the following
steps:
a. Press save/recall SETUP.
b. Press the main-menu Recall Factory Setup.
c. Press the side-menu OK Confirm Factory Init.
d. Wait for the Clock Icon to leave the screen.
e. CONFIRM the digitizing oscilloscope is setup as shown below.
Channel: CH1
Volt/div: 100 mV
Horizontal scale: 500 s/div
2. Set up digitizing oscilloscope for TV triggers by completing the following
steps:
a. Press TRIGGER MENU.
b. Press the main-menu Type pop-up until you select Video.
c. Press the main-menu Standard pop-up until you select 625/PAL.
d. Press the main-menu Line.
e. Use the keypad to set the line number to 7 (press 7, then ENTER).
f. Press VERTICAL MENU.