Functional Check
16
P7225 2.5 GHz 10X Active Probe User Manual
3. Using a ground lead and SMT KlipChip, connect the probe
ground to the instrument ground on the host instrument.
4. Using a standard tip, hold the probe to the center conductor of the
PROBE COMPENSATION output on the host instrument.
5. Press AUTOSET, or adj ust the host instrument to display a stable
calibration waveform. (You may need to adjust the probe offset to
display the waveform.)
NOTE. Now is a good time to perform a probe calibration routine
from your instrument. Refer to the online help of your host instrument
for calibration routine procedures.
6. Disconnect the probe tip from t he host instrument, and ground the
probe tip. (Connect the KlipChip to the probe tip.)
7. Set the probe offset to 0.0 V to m ake the host instrument display
at the ground reference.
8. Adjust the V/div to keep the waveform on the screen.