Configuration
18
P7225 2.5 GHz 10X Active Probe User Manual
6. Change the volts/division setting to the desired range, adjusting
the offset to keep the trace on the zero reference level.
NOTE.TheP7225 has a ±10.0 V offset range. The input volt age
linear operating range is ±4 V. See Figure 4.
If cursors are used on a TekConnect Interface host instrument, the
zero reference will be at the probe offset voltage.
0V
+14.0 V
--14.0 V
+4.0 V
-- 4 . 0 V
+10.0 V
--10.0 V
Nonoperating range (+30 V maximum nondestructive input voltage )
+10.0 V
--10.0 V
Nonoperating range (--30 V maximum nondestructive input voltage )
Maximum offset rangeMaximum AC signal amplitude
Figure 4: Dynamic and offset limitations