Commands Listed in Alphabetical Order
MEASUrement:
CLEARSNapshot (No Query Form)
Removes the measurement snapshot display.
Group
Measurement
Syntax
MEASUrement:CLEARSNapshot
Related Commands
CLEARMenu
MEASUrement:GATing
Species or returns the measurement gating setting.
Group
Measurement
Syntax
MEASUrement:GATing {OFF|SCREen|CURS or}
MEASUrement:GATing?
Arguments
OFF turns off measurement gating (full record).
SCREen turns o n gating, using the left and right edges of the screen.
CURSor limits measurements to the portion of the waveform between the vertical
bar cursors, even if they are off screen.
Examples
MEASUREMENT:GATING CURSOR turns on measurement gating using the cursors
as limits.
MEASUREMENT:GATING? might return :ME ASUREMENT:GATING CURSOR
indicating that measurements are limitedtotheportionofthewaveformbetween
the vertical bar cursors.
MEASUrement:IMMed? (Query Only)
Returns all immediate measurement set up parameters.
Group
Measurement
Syntax
MEASUrement:IMMed?
2-206 MSO4000 and DPO4000 Series Programmer Manual