Commands Listed in Alphabetical Order
Group
Power
Syntax
POWer:SWLoss:REFLevel:PER Cent:LOWCurrent <NR3>
POWer:SWLoss:REFLevel:PER Cent:LOWCurrent?
Arguments
<NR3> is the low voltage reference level percent.
Examples
POWER:SWLOSS:REFLEVEL:PER CENT:LOWCURRENT 15 sets the reference low
current to 15%.
POWER:SWLOSS:REFLEVEL:PER CENT:LOWCURRENT ? might return
:POWER:SWLOSS:REFLEVEL:PE RCENT:LOWCURRENT 5.0000 indicating that
the reference low current is set to 5%.
POWer:SWLoss:REFLevel:PERCent:LOWVoltage
Sets or returns the low voltage reference level used in switching loss power
measurements in percent.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:SWLoss:REFLevel:PER Cent:LOWVoltage <NR3>
POWer:SWLoss:REFLevel:PER Cent:LOWVoltage?
Arguments
<NR3> is the low voltage reference level in percent.
Examples
POWER:SWLOSS:REFLEVEL:PER CENT:LOWVOLTAGE 10 sets the reference low
voltage to 10%.
POWER:SWLOSS:REFLEVEL:PER CENT:LOWVOLTAGE ? might return
:POWER:SWLOSS:REFLEVEL:PE RCENT:LOWVOLTAGE 5.0000 indicating that
the reference low voltage is set to 5%.
POWer:SWLoss:TO FF:ENERGY:MAX? (Query Only)
Returns the maximum Toff energy for the switching loss calculation.
Conditions
This command requires a DPO4PWR application module.
2-306 MSO4000 and DPO4000 Series Programmer Manual