Commands Listed in Alphabetical Order
Syntax
POWer:HARMoni
cs:RESults:HAR<1-400>:RMS:AB Solute?
Examples
POWER:HARMONICS:RESULTS:HA R400:RMS:ABSOLUTE? might return
:POWER:HARMONICS:RESULTS:H AR400:RMS:ABSOLUTE 1.0 indicating that
the RMS magnitude of the harmonic is set to 1.0 A.
POWer:HARMonics:RESults:HAR<1-400>:RMS:PERCent? (Query Only)
Returns the RMS magnitude of the harmonic expressed as a percentage of the
fundamental.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:HARMonics:RESults:HA R<1-400>:RMS:PERCent?
Examples
POWER:HARMONICS:RESULTS:HA R400:RMS:PERCENT? might return
:POWER:HARMONICS:RESULTS:H AR400:RMS:PERCENT 33 indicating that the
RMS magnitude of the harmonic is set to 33%.
POWer:HARMonics:RESults:HAR<1-400>:TEST:IEC:CLASSALIMit? (Query On
ly)
Species if the IEC Class A higher harmonic limit (and conditions) are met.
NOTE. The command returns NA if the standard does not specify a limit or the
equipment is not class A.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:HARMonics:RESults:HA R<1-400>:TEST:IEC:CLASSALIMit?
Returns
PASS, FAIL,orNA.
MSO4000 and DPO4000 Series Programmer Manual 2-261