Commands Listed in Alphabetical Order
POWer:HARMon
ics:RESults:HAR<1-400>:TEST:IEC:NORMAL? (Query Only)
Specifies if the Normal IEC harmonic limits are met.
NOTE. The command returns NA if the standard does not specify a limit.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:HARMonics:RESults:H AR<1-400>:TEST:IEC:NORMAL?
Returns
PASS, FAIL,orNA.
POWer:HARMonics:RESults:HAR<1-400>:TEST:IEC:POHCLIMit? (Query O nly)
Specifies if the higher harmonic limit (and conditions) for the 21st and higher
order odd harmonics are met.
NOTE. The command returns NA if the limit does not apply for a specific
harmonic.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:HARMonics:RESults:H AR<1-400>:TEST:IEC:POHCLIMit?
Returns
PASS, FAIL,orNA.
POWer:HARMonics:RESults:HAR<1-400>:TEST:MIL:NORMAL? (Query Only)
Returns the test result for the specified harmonic for the MIL-STD-1399 testing
standard.
This query is analogous to that for the IEC 61000-3-2 standard
POWer:HARMonics:RESults:HAR<1-400>:TEST:IEC:NORMAL? command.
2-262 MSO4000 and DPO4000 Series Programmer Manual