Commands Listed in Alphabetical Order
Related Commands
MEASUrement:R
EFLevel:METHod
Arguments
<NR3> is the mid reference level, in volts. The default is 0.0 V.
Examples
MEASUREMENT:REFLEVEL:ABSO LUTE:MID2 0.5 sets the mid reference level for
the d elay waveform to
0.5 V.
MEASUREMENT:REFLEVEL:ABSO LUTE:MID2? might return
:MEASUREMENT:REFL
EVEL:ABSOLUTE:MID2 0.5000E+00
indicating that the
absolute mid2 reference level is set to 0.5 V.
MEASUrement:REFLevel:ABSolute:MID<x>
Sets or returns the mid reference level for channel <x>, where x is the
measurement channel.
Group
Measurement
Syntax
MEASUrement:REFLevel:ABSo lute:MID<x> <NR3>
MEASUrement:REFLevel:ABSo lute:MID<x>?
Arguments
<NR3> is the mid reference level in volts.
MEASUrement:REFLevel:METHod
Species or returns the reference level units used for measurement calculations.
NOTE. this command affects the associated reference level parameter
for all MEASurements:IMMed and t
he eight periodic measurements.
To change the parameter for individual measurements, use the
MEASUrement:MEAS<x>:REFLevel commands.
Group
Measurement
Syntax
MEASUrement:REFLevel:METH od {ABSolute|PERCent}
MEASUrement:REFLevel:METH od?
Arguments
ABSolute species that the reference levels are set e
xplicitly using the
MEASUrement:REFLevel:ABSolute commands. This method is useful when
2-230 MSO4000 and DPO4000 Series Programmer Manual