Tektronix DPO4000 Manuel d'utilisateur Page 224

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 585
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 223
Commands Listed in Alphabetical Order
MEASUrement:
IMMed:VALue? (Query Only)
Returns the value of the measurement specied by the MEASUrement:IMMed:
TYPe command. The measurement is immediately taken on the source(s)
specied by a
MEASUrement:IMMed:SOUrce command.
NOTE. A change to HORizontal:MAIn:SCALe or CH<x>:SCALe will not
necessarily have taken affect if immediately followed by this command.
Group
Measurement
Syntax
MEASUrement:IMMed:VALue?
Related Commands
MEASUrement:IMMed:TYPe, MEASUrement:IMMed:SOUrce, *ESR?,
ALLEv?
Examples
MEASUREMENT:IMMED:VALUE? might return : MEASUREMENT:IMMED:VALUE
9.9000E+37.
If the measurement has an error or warning associated with it,
then an item is a dded to the error queue. The error can be checked for with the
*ESR? and ALLEv? commands.
MEASUrement:INDICators? (Query Only)
Returns all measurement indicator parameters.
Group
Measurement
Syntax
MEASUrement:INDICators?
Examples
MEASUREMENT
:INDICATORS?
might return
MEASUREMENT:INDICATORS:ST ATE MEAS1;NUMHORZ 0;NUMVERT 4;HORZ1
7.5E0;HORZ2 -3.400000095367E0;HORZ3 0.0E0;HORZ4 0.0E0;VERT1
-6.351123E-6;VERT2 -3.179753E-6;VER T3 -6.40943E-6;VERT4
-6.403E-6
MEASUrement:INDICators:HORZ<x>? (Query Only)
Returns t
he position of the specied horizontal measurement indicator < x>, where
<x> can be 1, 2, 3, or 4.
2-214 MSO4000 and DPO4000 Series Programmer Manual